Welcome to Nuclear Fusion and Plasma Physics, Today is Share:

NUCLEAR FUSION AND PLASMA PHYSICS ›› 2020, Vol. 40 ›› Issue (3): 200-207.DOI: 10.16568/j.0254-6086.202003002

• Plasma Physics • Previous Articles     Next Articles

Development and test of the beam emission spectroscopy diagnostic based on neutral beam injection on HL-2A tokamak

WU Yi-fan1, 2, YU Yi1, 2, KE Rui2, XU Min2, NIE Lin2, WU Ting2, SUN Ai-ping2, LAN Tao1, YUAN Bo-da1, 2, LIU Hao1, 2, GONG Shao-bo1, 2, LONG Ting2, DUAN Xu-ru2, YE Min-you1, HL-2A team2   

  1. (1. School of Physics, University of Science and Technology of China, Hefei 230026; 2. Southwestern Institute of Physics, Chengdu 610041)
  • Received:2019-04-04 Revised:2020-04-23 Online:2020-09-15 Published:2020-09-16

HL-2A装置中性束发射光谱 诊断系统的研制与测试

吴一帆1, 2,余 羿1, 2,柯 锐2,许 敏2,聂 林2, 吴 婷2,孙爱萍2,兰 涛1,袁博达1, 2,刘 灏1, 2, 龚少博1, 2,龙 婷2,段旭如2,叶民友1,HL-2A团队2   

  1. (1. 中国科学技术大学物理学院,合肥 230026;2. 核工业西南物理研究院,成都 610041)
  • 作者简介:吴一帆(1991-),男,安徽池州人,博士研究生,从事等离子体物理诊断及实验研究。
  • 基金资助:
    国家重点研发计划专项(2017YFE0300405);国家磁约束核聚变能发展研究专项(2015GB120002,2015GB104000);国家自然科学基金(11875124,U1867222,11575055,11705052)

Abstract: A 32-channels beam emission spectroscopy (BES) diagnostic system is developed and tested on HL-2A tokamak, which enables the 2D measurements of the long wavelength () electron density fluctuation with both high time and spatial resolutions () in a 2D field zone from radial position r=12~44cm and poloidal position -7.5~+7.5cm. The diagnostic consists of the sets of asymmetric lenses in the vacuum, fibers, high-efficiency detector modules and attached cooler and vacuum pump. In the low frequency (f<100kHz), the photon noise makes the main contribution to the noise level of the whole system, while the e-noise catches the similar level with the photon noise in the higher frequency. During the typical discharge of HL-2A, the signal to noise ratio (SNR) of the whole system is above 3 for the turbulence below 200kHz frequency.

Key words: HL-2A tokamak, Beam emission spectroscopy, Electron density fluctuation, 2D turbulence measurement

摘要: 在HL-2A装置上完成了一套32通道束发射诊断系统(BES),可对径向r=12~44cm, 极向-7.5~+7.5cm二维空间范围内的长波长()电子密度扰动信息进行测量,其时间分辨率达到0.5ms,空间分辨率1~2cm。系统由内置于真空室的非对称镜头组、传输光纤、高性能探测器模块以及辅助的冷却和真空设备构成。系统的噪声在低频时(f<100kHz)主要由散粒噪声贡献,在较高频率时由散粒噪声和e噪声共同决定。在典型的HL-2A装置放电模式中,对于200kHz以下的扰动,该系统的信噪比(SNR)均大于3。

关键词: HL-2A装置, 束发射光谱, 密度涨落, 二维湍流测量

CLC Number: